| Model Number |
CR2 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that the magnet was missing from the lid of their device.This will result in an inability of the device to detect the lid being opened or closed.In this state, a delay in defibrillation may occur, as the user has to push the power button underneath the lid to turn the device on.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer will receive a replacement lid and battery.The device was not returned to stryker for investigation.A cause of the reported issue could not be determined.
|
| |
|