| Model Number |
AA10 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3 product analysis:evaluation determined that bearing detached.The likely cause of failure is corrosion.It was also noted difficult to insert tool.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Repair request initiated for device with the report of difficulty with assembly.It was reported there was no patient involvement.Repair request escalated to a product event due to evaluation finding of bearing inside the tube was damaged.
|
| |
|
Search Alerts/Recalls
|