• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARD® EAGLE® INFLATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARD® EAGLE® INFLATION DEVICE Back to Search Results
Catalog Number 325130
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the erics procedure the balloon dilation catheter failed to inflate properly because contrast liquid leakage from the manometer during fluid insufflation.(batch# bmgwat01), (batch# bmgyfm44).
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during the erics procedure the balloon inflation device failed to inflate properly because contrast liquid leakage from the manometer during fluid insufflation.(batch#bmgwat01, batch#bmgyfm44).
 
Additional Manufacturer Narrative
The reported event is confirmed cause unknown.Visual evaluation noted received one photo sample that shows side profile of contrast solution leaking out of eagle inflation device.Based on the photo sample received the reported event is confirmed.No actions can be taken at this time since a root cause was not identified.Dhr review did not show any problems or conditions that contributed to the reported event.The instructions for use were found adequate and state the following: description: the bard eagle inflation device is a one-piece, plastic, 10cc disposable inflation device with a lock lever design that controls the piston, a pressure gauge, and a high-pressure connecting tube with a male rotating adapter.The pressure gauge measures pressures ranging from vacuum to 30 atm; the gauge is marked in 1 atm increments.The gauge also has an inner scale of comparable psi measurements.The accuracy of the pressure gauge has been determined to be within 1 atm over the range.Indications: the inflation device is recommended for use while performing urological balloon dilation procedures to inflate the balloon, sustain pressure, monitor pressure within the balloon, and deflate the balloon.Contraindication: none.Warnings: use only liquid inflation media.Do not inflate with air.Always follow the manufacturers directions accompanying the balloon dilation catheter for instructions for use, maximum balloon inflation pressure, precautions, and warnings for that device.This is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable regulations.Precaution: before use, inspect the device to verify that no damage has occurred during shipping and handling.Caution: federal (usa) law restricts this device to sale by or on the order of a physician.Instructions for use: preparation: make all aspiration and injection maneuvers with the lock lever pushed left, i.E., unlocked.Unlock the piston by pushing the lock lever left.In this position, you can freely pull the piston back for aspiration, or push it forward for injection.To lock the piston in position, slide the lever right to the straight up position.1.Prepare a solution of contrast medium and normal saline in a small sterile bowl or in the well provided with the package.Check balloon catheter and contrast medium instructions for specific contrast mixture recommendations.2.Orient the tubing downward into the contrast medium.3.Push the release lever left and aspirate enough solution to fill the syringe.4.Hold the device upright to purge the air from the syringe and connecting tube.Tap the syringe lightly, if necessary, to remove all the air bubbles and fill the connecting tube completely.5.Inspect the syringe and tubing to ensure that the device has been completely purged.6.Adjust the syringe volume to 3 to 4cc.If more contrast solution is needed, submerge the syringe tip into the basin of solution and aspirate.Attaching the inflation device to the balloon dilation catheter: 1.Prepare and test the balloon dilation catheter according to the manufacturers directions for use.2.If a separate syringe was used to prepare the balloon catheter, remove it.Create a fluid-fluid connection between the balloon and the connecting tube (male rotating adapter) of the inflation device by placing a drop of contrast solution from the syringe into each hub.3.Hand tighten the hubs securely.Balloon inflation and deflation: 1.Release the lock lever and allow the piston to move forward into neutral position (0 atm).2.To inflate the balloon, engage the lock lever, turn the palm grip on the piston clockwise slowly until the desired inflation pressure is reached.The lock lever maintains the increasing pressure.3.To deflate the balloon, push the lock lever left, releasing the piston, and pull back.Slide the lock lever back to lock, if desired.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® EAGLE® INFLATION DEVICE
Common Device Name
INFLATION DEVICE
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key20292481
Report Number1018233-2024-06016
Device Sequence Number8012456
Product Code KOE
UDI-Device Identifier00801741067259
UDI-Public(01)00801741067259(11)221005(17)250831(10)BMGWAT01
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K962611
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Catalogue Number325130
Device Lot NumberBMGWAT01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2024
Supplement Date Received by Manufacturer01/22/2025
Initial Report FDA Received Date09/24/2024
Supplement Report FDA Received Date01/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-