| Model Number |
DSX200H11C |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of particles in the air path.The patient reports seeing the particles in 2023 but could not recall the month.The particles were seen in the pipe and hose.The patient reports cleaning the device with warm soap and water every 3-4 weeks.The patient no longer uses the device and will return it shortly.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of particles in the air path.The patient reports seeing the particles in 2023 but could not recall the month.The particles were seen in the pipe and hose.The patient reports cleaning the device with warm soap and water every 3-4 weeks.The patient no longer uses the device and will return it shortly.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.This report is being submitted due to previous report stating a follow up report will be submitted when the manufacturer's investigation is complete.No device has been returned.No further evaluation is possible at this time.Attempts to have the device returned for evaluation is possible at this time.Attempts to have the device returned for evaluation and investigation were unsuccessful.A final report is being submitted.If new information becomes available that changes the outcome of the investigation and associated medical device reporting a follow up/supplemental report will be completed.
|
| |
|
Search Alerts/Recalls
|