ABBOTT DIABETES CARE LTD FREESTYLE LIBRE 3; CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 72075-01
Medical Device Problem Codes Device Alarm System (1012); Application Program Problem (2880)
Health Effect - Clinical Codes Hypoglycemia (1912); Diaphoresis (2452)
Date of Event 09/07/2024
Type of Reportable Event Serious Injury
Event or Problem Description
An alarm issue was reported with the abbott diabetes care (adc) device in use with samsung phone with android operating system version 14.The customer experienced a signal loss and was unable to receive glucose alarms.As a result, the customer was not alerted of changes in glucose level and experienced sweating and was unable to self-treat, requiring treatment of an "okluko" injection by healthcare professional (friend).There was no report of death or permanent impairment associated with this event.
 
Additional Manufacturer Narrative
Investigation is pending at this time.A follow up will be submitted once additional information is obtained.The device manufacturing date does not apply.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Additional Manufacturer Narrative
An extended investigation has been performed for the reported complaint.The customer reported for signal loss.Sensor was inserted in the simvivo test fixture.The app was downloaded and initial app setup was completed.Sensor was scanned by the app and enabled ¿signal loss alarm¿ feature in the app and placed device in faraday bag to interrupt signal to sensor.Removed device from bag and confirmed sensor was reconnected within few minutes.Alarms functioned as intended.Attempted to replicate customer complaint using similar configuration.However, the reported issue could not be replicated and the system functioned as intended under the tested conditions.Based on the customer¿s reported application, potential causes and device history were also examined, but no issues were identified during replication that could have led to the reported issue.Therefore, the issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
An alarm issue was reported with the abbott diabetes care (adc) device in use with samsung phone with android operating system version 14, app version 3.5.1.10363.The customer experienced a signal loss and was unable to receive glucose alarms.As a result, the customer was not alerted of changes in glucose level and experienced sweating and was unable to self-treat, requiring treatment of an "okluko" injection by healthcare professional (friend).There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 3
Common Device Name
CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE LTD
range road
witney 12345 -700
UK  12345-7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key20295810
Report Number2954323-2024-37607
Device Sequence Number7284337
Product Code QLG
UDI-Device Identifier00357599816001
UDI-Public00357599816001
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K213996
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number72075-01
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2024
Supplement Date Received by Manufacturer11/08/2024
Initial Report FDA Received Date09/24/2024
Supplement Report FDA Received Date11/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/13/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-