| Model Number |
72075-01 |
| Medical Device Problem Codes |
Device Alarm System (1012); Application Program Problem (2880)
|
| Health Effect - Clinical Codes |
Hypoglycemia (1912); Diaphoresis (2452)
|
| Date of Event |
09/07/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
An alarm issue was reported with the abbott diabetes care (adc) device in use with samsung phone with android operating system version 14.The customer experienced a signal loss and was unable to receive glucose alarms.As a result, the customer was not alerted of changes in glucose level and experienced sweating and was unable to self-treat, requiring treatment of an "okluko" injection by healthcare professional (friend).There was no report of death or permanent impairment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is pending at this time.A follow up will be submitted once additional information is obtained.The device manufacturing date does not apply.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
An extended investigation has been performed for the reported complaint.The customer reported for signal loss.Sensor was inserted in the simvivo test fixture.The app was downloaded and initial app setup was completed.Sensor was scanned by the app and enabled ¿signal loss alarm¿ feature in the app and placed device in faraday bag to interrupt signal to sensor.Removed device from bag and confirmed sensor was reconnected within few minutes.Alarms functioned as intended.Attempted to replicate customer complaint using similar configuration.However, the reported issue could not be replicated and the system functioned as intended under the tested conditions.Based on the customer¿s reported application, potential causes and device history were also examined, but no issues were identified during replication that could have led to the reported issue.Therefore, the issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
An alarm issue was reported with the abbott diabetes care (adc) device in use with samsung phone with android operating system version 14, app version 3.5.1.10363.The customer experienced a signal loss and was unable to receive glucose alarms.As a result, the customer was not alerted of changes in glucose level and experienced sweating and was unable to self-treat, requiring treatment of an "okluko" injection by healthcare professional (friend).There was no report of death or permanent impairment associated with this event.
|
| |
|
Search Alerts/Recalls
|