SMITH & NEPHEW, INC. PROBE SAPHYRE 90DEG; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
| Catalog Number |
7209686 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during a joint cavity debridement when the probe saphyre wand was used for the first time, the insulating layer was split inside the articular cavity, the wand was quickly removed from the patient.The procedure was completed using an s+n backup device.There was no surgical delay, and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h3, h6: the reported device was received for evaluation.Visual inspection observed the tip is worn from use and the shaft insulation is damaged from use.The shaft insulation is split near the tip of the device exposing the internal shaft metal.Functional evaluation revealed the unit had no issues activating in use.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences.Other factors include excessive force, using the probe as a lever in the joint space or contact with another instrument during the procedure.Based on this investigation, there is no evidence to suggest a design, material, or manufacturing issue.Therefore, the need for corrective action is not indicated.
|
| |
|