| Model Number |
004628-005 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product was returned for evaluation.Should new relevant information become available, a supplemental report will be submitted.Only use u-100 humalog or u-100 novolog with your pump.Only u-100 humalog and u-100 novolog have been tested and found to be compatible for use in the pump.Use of insulin with greater or lesser concentration can result in an over delivery or under delivery of insulin.This can cause hypoglycemia (low bg) or hyperglycemia (high bg) events.
|
| |
|
Event or Problem Description
|
|
It was reported that occlusion alarms occurred.Customer declined follow up from tandem technical support.Reportedly the customer is using u-500 insulin.There was no reported adverse impact.No additional information was provided.
|
| |
|