• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP Back to Search Results
Catalog Number 1801-58
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that four syntel silicone embolectomy catheter - regular tip balloon ruptured during use.No injury was reported to the patients.More information in regard to the reported incidents were requested.Note: this is report 1 of 4 related to the first catheter used in the event.Reports 1220948-2024-00188, 1220948-2024-00189, and 1220948-2024-00190 were submitted for the second, third, and forth devices involved in this event.
 
Additional Manufacturer Narrative
We have not received the device for evaluation.Therefore, we could not conclusively determine the root cause of the incident.Balloon rupture is a known complication with this type of product and is not an indication of a systemic issue with the product.As stated in the ifu: in common with other catheterization procedures, complications may occur.These may include intimal disruption, vessel wall perforation, balloon rupture with fragmentation, tip separation, local or systemic infection, arterial thrombosis, local hematomas, air embolus, plaque, arterial dissection, and hemorrhage.Exposure to calcified plaques may increase the risk of balloon rupture.The lot history record for the device was reviewed, no issues were found during manufacturing or packaging that would cause or contribute to the reported event.We have not received any similar complaints related to this lot number.Note: this is report 1 of 4 related to the first catheter used in the event.Reports 1220948-2024-00188, 1220948-2024-00189, and 1220948-2024-00190 were submitted for the second, third, and fourth devices involved in this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNTEL SILICONE EMBOLECTOMY CATHETER - REGULAR TIP
Common Device Name
EMBOLECTOMY CATHETER - REGULAR TIP
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC
63 second ave
burlington MA 01803
Manufacturer Contact
peter song
63 second ave
burlington, MA 01803
MDR Report Key20297838
Report Number1220948-2024-00187
Device Sequence Number7588727
Product Code DXE
UDI-Device Identifier00840663100941
UDI-Public(01)00840663100941(11)220407(17)270328(10)NSE2431(91)1801-58
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K992368
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 09/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1801-58
Device Lot NumberNSE2431
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2024
Initial Report FDA Received Date09/24/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-