• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL ET TUBE,HVT,7.5,NOVA PLUS; TUBE, TRACHEAL (W/WO CONNECTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL ET TUBE,HVT,7.5,NOVA PLUS; TUBE, TRACHEAL (W/WO CONNECTOR Back to Search Results
Catalog Number V5-10315
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that: "ett , leak detected.".
 
Additional Manufacturer Narrative
Qn# (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Additional Manufacturer Narrative
(b)(4).The customer returned one unit v5-10315 et tube,hvt,7.5,nova plus for investigation.The et tube was visually examined with and without magnification.Visual examination of the returned de-vice revealed that the sample appeared typical.No defects or anomalies were observed.Functional inspection was performed on the returned device to test for proper inflation and deflation.A lab inventory syringe was filled with air and attached to the pilot balloon.Upon injection of air, the balloon cuff was unable to stay inflated.The sample was submerged underwater to test for leaks.Upon inflation, the et tube leaked from a hole in the balloon cuff.A device history record review was performed with no evidence to suggest a manufacturing related cause.The ifu for this product was reviewed as a part of this complaint investigation.The ifu states, "the tracheal tube cuff inflation system, including the valve, should be checked prior to intubation.If a malfunction is a detected in any part of the system, the tube should not be used." "care should be taken to avoid damaging the thin-walled cuffs during intubation.If cuff is damaged, the tube should not be used." the ifu also states, "the tracheal tube cuff should be slowly inflated with the minimum amount of air required to provide an effective tracheal seal.Do not inflate with a measured volume of air or by "feel" of pressure in the syringe since little resistance should be felt during inflation.Cuff volume and pressure should be monitored routinely to assure that an adequate tracheal seal is maintained and that the cuff has not become overinflated." the reported complaint of "leak - device leak - tube" was confirmed based upon the sample received.The returned et tube was found to have a hole in the balloon cuff which prevented it from being able to stay inflated.A device history record review was performed with no evidence to suggest a manufacturing related cause.All et tubes are 100% inspected for inflation and deflation at the time of manufacturing.It is unlikely that this type of damage was present at the time of manufacturing.Therefore, based upon the damage observed, unintentional user error caused or contributed to this event.
 
Event or Problem Description
It was reported that: "ett , leak detected".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ET TUBE,HVT,7.5,NOVA PLUS
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
sachin sony
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key20298877
Report Number3003898360-2024-01177
Device Sequence Number7792399
Product Code BTR
UDI-Device Identifier04026704588787
UDI-Public(01)04026704588787(11)240121(17)290120(10)73A2400515(21)
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K822082
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberV5-10315
Device Lot Number73A2400515
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2024
Supplement Date Received by Manufacturer10/29/2024
Initial Report FDA Received Date09/25/2024
Supplement Report FDA Received Date11/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED.; NOT REPORTED.
Patient SexUnknown
-
-