• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 24MM +2 LAT BASEPLATE, MODULAR; SHOULDR PROSTH, REVERSE CONFIG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 24MM +2 LAT BASEPLATE, MODULAR; SHOULDR PROSTH, REVERSE CONFIG Back to Search Results
Model Number 24MM +2 LAT BASEPLATE, MODULAR
Medical Device Problem Codes Break (1069); Defective Device (2588)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/12/2024
Type of Reportable Event Malfunction
Event or Problem Description
Additional information received on 9/20/2024: this was discovered during an rtsa procedure.The sales representative reported that the surgeon attempted to screw in the baseplate when the morse taper came apart, and the central screw fell off the baseplate.All products were retrieved.The case was completed using a new ar-9561-25s univers revers modular central screw and an ar-9560-24-2 arthrex univers revers modulas baseplate.The case was not delayed, and additional anesthesia was not needed.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On (b)(6) 2024, it was reported by a sales representative via email that an ar-9561-25s univers revers modular central screw and an ar-9560-24-2 arthrex univers revers modulas baseplate failed.This was discovered during a procedure on (b)(6) 2024.
 
Additional Manufacturer Narrative
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error due to misaligned insertion and/or improper bone preparation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
24MM +2 LAT BASEPLATE, MODULAR
Common Device Name
SHOULDR PROSTH, REVERSE CONFIG
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key20300753
Report Number1220246-2024-07836
Device Sequence Number7726510
Product Code PHX
UDI-Device Identifier00888867296497
UDI-Public(01)00888867296497(11)240731(17)290630(10)15295273
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K173900
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24MM +2 LAT BASEPLATE, MODULAR
Device Catalogue NumberAR-9560-24-2
Device Lot Number15295273
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2024
Supplement Date Received by ManufacturerNot provided
06/10/2025
Initial Report FDA Received Date09/25/2024
Supplement Report FDA Received Date09/30/2024
07/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/31/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-