BOLTON MEDICAL, INC. RELAY PRO NBS THORACIC STENT-GRAFT SYSTEM; STENT, ENDOVASCULAR GRAFT, AORTIC
|
Back to Search Results |
|
| Catalog Number |
28-N4-32-104-32U |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Failure to Seal (4070); Device Endoleak (4074)
|
| Health Effect - Clinical Codes |
Hemoptysis (1887); Aneurysm expansion (4597); Clinical Endoleak (4843)
|
| Date of Event |
08/25/2024
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
"subject presented a penetrating aortic ulcer (pau) and was treated with relaypro nbs on (b)(6) 2024.Subject was discharged to an outpatient facility on (b)(6) 2024 and unfortunately passed away on (b)(6) 2024; an autopsy was performed.Pi has declared the death undetermined if related to procedure or device.Imaging has been requested.As per death documentation: cta chest - aorta: redemonstration of a patient endograft repair of a penetrating descending thoracic aortic ulcer/aneurysm.There is new opacification of the aneurysm sac posterior to the abdominal aorta concerning for an endoleak, likely type i or type iv.The aneurysm sac measures up to 4.5 cm in diameter.There remains some surrounding hypodensity that is unchanged compared to prior favored to represent an evolving old blood products." patient outcome: "patient death".
|
| |
|
Additional Manufacturer Narrative
|
|
Bolton medical, inc.(d/b/a terumo aortic), herein known as the "company", is submitting this report pursuant to 21 cfr part 803, has made reasonable efforts to obtain complete information, and has provided as much as is available to the company as of the submission date of this report.This report is based on information obtained by the company, which may not have been able to fully investigate or verify prior to the date the report was required by the regulations.Any required fields that are unpopulated are blank because the information is currently unknown, unavailable, or not applicable.This report does not constitute an admission or a conclusion by anyone that the device caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the certain regulations, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
"subject presented a penetrating aortic ulcer (pau) and was treated with relaypro nbs on (b)(6) 2024.Subject was discharged to an outpatient facility on (b)(6) 2024 and unfortunately passed away on (b)(6) 2024; an autopsy was performed.Pi has declared the death undetermined if related to procedure or device.Imaging has been requested.As per death documentation: cta chest - aorta: redemonstration of a patient endograft repair of a penetrating descending thoracic aortic ulcer/aneurysm.There is new opacification of the aneurysm sac posterior to the abdominal aorta concerning for an endoleak, likely type i or type iv.The aneurysm sac measures up to 4.5 cm in diameter.There remains some surrounding hypodensity that is unchanged compared to prior favored to represent an evolving old blood products." patient outcome: "patient death".
|
| |
|
Additional Manufacturer Narrative
|
|
Bolton medical, inc.(d/b/a terumo aortic), herein known as the "company", is submitting this report pursuant to 21 cfr part 803, has made reasonable efforts to obtain complete information, and has provided as much as is available to the company as of the submission date of this report.This report is based on information obtained by the company, which may not have been able to fully investigate or verify prior to the date the report was required by the regulations.Any required fields that are unpopulated are blank because the information is currently unknown, unavailable, or not applicable.This report does not constitute an admission or a conclusion by anyone that the device caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the certain regulations, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Search Alerts/Recalls
|
|
|