| Model Number |
25MM CENTRAL SCREW, MODULAR |
| Medical Device Problem Codes |
Defective Device (2588); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 9/17/2024, it was reported by a sales representative via email that an ar-9561-25s univers revers modular central screw and an ar-9560-24-2 arthrex univers revers modulas baseplate failed.This was discovered during a procedure on (b)(6) 2024.
|
| |
|
Event or Problem Description
|
|
Additional information received on 9/20/2024: this was discovered during an rtsa procedure.The sales representative reported that the surgeon attempted to screw in the baseplate when the morse taper came apart, and the central screw fell off the baseplate.All products were retrieved.The case was completed using a new ar-9561-25s univers revers modular central screw and an ar-9560-24-2 arthrex univers revers modulas baseplate.The case was not delayed, and additional anesthesia was not needed.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error due to misaligned insertion and/or improper bone preparation.
|
| |
|
Search Alerts/Recalls
|