| Model Number |
OTV-S7H-NA-12E |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the subject device had blurry blue screen.The issue was found during reprocessing.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: a leak from cable.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: blurry blue screen" was due to scratches on bad ccd lens and a leak from cable due to tiny pin hole on cable.The cause was traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was reported that the subject device had blurry blue screen.The issue was found during reprocessing.There was no patient involvement.
|
| |
|