| Catalog Number |
00590103300 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an initial total knee arthroplasty, the tibial impactor head fractured while impacting the tibial component.No backup instrumentation was needed as the device was sufficiently impacted.There was no patient impact and no adverse events have been reported as a result of the malfunction.All available information has been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: medical product: femoral cr impactor pad: catalog #42509909400, lot#64337660.The product has been received by zimmer biomet and the investigation is in process.Upon completion of the investigation, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; g3; h2; h3; h6; h10; h11.Visual examination of the returned product identified signs of use (nicked or gouged) and is confirmed fractured.All pieces were returned.Device history record was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.Complaint sample was evaluated and the reported event was confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information at time of this report.
|
| |
|