• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-190-10-3D
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was observed that during the device inspection, the flex deflectable videoscope exhibited a deterioration of the glues on the distal end.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.Based on the results of the investigation, and although we could not specify the root cause of the suggested event, it is likely the defect was caused by occurred by physical stress by dropping the distal end, chemical stress from used chemical solutions, or the like.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The event can be prevented by following the instructions for use ltf-190-10-3d operation manual which state: important information ¿ please read before use: contraindications, warnings, and precautions: do not strike, hit, or drop the endoscope¿s distal end, rigid portion, bending section, control section, universal cord, video connectors, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, rigid portion, bending section, control section, universal cord, video connectors, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.The event can be prevented by following the instructions for use ltf-190-10-3d reprocessing manual 3.1 which state: compatibility summary: methods listed as compatible in table 3.1 and 3.2 are compatible for routine use only when used according to manufacturer¿s instructions.Repeated use and reprocessing of endoscopes and accessories lead to gradual wear and tear.Furthermore, reprocessing methods that employ higher temperatures and more caustic/corrosive materials may lead to faster deterioration.In general, sterilization processes are harsher on equipment than disinfection processes.Before each patient procedure, inspect the endoscope and accessories for damage, according to the instructions described in this manual and its companion operation manual.Reprocessing manual: 3.1 compatibility summary: list of compatible methods validated in terms of material durability: sterrad® 50/100s/200/nx¿: sterilization by sterrad® 50/100s/200/nx¿ system may deteriorate the adhesive of the insertion section.Depending on the circumstances, replacement of the insertion section may be required.Before use, confirm that the endoscope is free from any damage or other irregularities.For further details, contact olympus.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE
Common Device Name
FLEX 3D DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20301312
Report Number9610595-2024-19175
Device Sequence Number12978400
Product Code GCJ
UDI-Device Identifier04953170434938
UDI-Public04953170434938
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K123365
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberLTF-190-10-3D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2024
Supplement Date Received by Manufacturer10/16/2024
Initial Report FDA Received Date09/25/2024
Supplement Report FDA Received Date10/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-