MAUDE Adverse Event Report: EXACTECH, INC. LOGIC CR FEMORAL CEM, LEFT SZ 2.5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC CR FEMORAL CEM, LEFT SZ 2.5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-010-03-0225
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/21/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(6) -2024-00850 d10 concomitant devices (b)(6) - 02-012-49-2509 - logic tibial insert slope ++, sz 2.5, 9 mm (b)(6) - 02-012-45-2525 - lgc tibial fit tray cem sz 2.5f / 5t (b)(6) - 200-02-35 - three peg patella 35mm (b)(6) - 201-78-15 - holding pin mini sharp point 4 pk (b)(6) - 201-78-81 - 3 trocar, mod.Hex 2pk (b)(6) - 21-78-23 - threaded pin size 2.3 collared (b)(6) - 521-78-23 - threaded pin size 2.3 collared (b)(6) - 521-78-32 - threaded pin size 3.0 collarless (b)(6) - a10012 - gps implant kit v2.Manufacturer narrative: the revision reported in was most likely the result of a combination of prosthesis wear secondary to contributions from implant alignment and/or instability of the knee as well as loosening due to mechanical loss of fixation at the cement/implant interface.However, this cannot be confirmed as the devices were not returned for evaluation and the loosening could not be confirmed on the pre-revision radiograph.
 
Event or Problem Description
It was reported that a 63yo male patient, initial left knee implanted on (b)(6) 2019, underwent a revision procedure on (b)(6) 2024, approximately 5 years post the initial procedure.The patient was revised due to femur loosening and poly wear.- everything was removed and exchanged to a competitor¿s product.There were no surgical delays or device breakages during the procedure.X-rays were provided.The patient was last known to be in stable condition following the event.No explanted devices are available for analysis.They were disposed of by the hospital.Device images were provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC CR FEMORAL CEM, LEFT SZ 2.5
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key20301350
Report Number1038671-2024-03696
Device Sequence Number12982559
Product Code JWH
UDI-Device Identifier10885862158000
UDI-Public10885862158000
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K111400
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/11/2018
Device Catalogue Number02-010-03-0225
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2024
Initial Report FDA Received Date09/25/2024
Date Device Manufactured09/13/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age63 YR
Patient SexMale
Patient Weight79 KG
-
-