ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 10" (25 CM) APPX 1.0 ML, SMALLBORE EXT SET W/6-PORT NANOCLAVE¿ MANIFOLD, CHECK V; STOPCOCK, I.V. SET
|
Back to Search Results |
|
| Catalog Number |
AM6154 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The event involved a 10" (25 cm) appx 1.0 ml, smallbore ext set w/6-port nanoclave¿ manifold, check valve (orange, glow, blue, purple, yellow, green rings), nanoclave¿, luer lock where the customer reported a leak from the tubing bonding connection point.There was patient involvement but no harm or adverse event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.The investigation is pending.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: the investigation found additional malfunctions, which makes this report 1 of 4.Investigation summary: received five (5) used.List #am6154, 10" (25 cm) appx 1.0 ml, smallbore ext set w/6-port nanoclave¿ manifold, check valve (orange, glow, blue, purple, yellow, green rings), nanoclave¿, luer lock; lot #13907590.As received, sample #3 observed to have the tubing separated from the manifold.Sample #4 observed to have the nanoclave body separated from the spike.No other physical damage or anomalies observed.Sample #2 was unable to confirm complaint.The other 2 samples, confirmed customer complaint of leaking from tubing bonding connection point.Probable cause, incomplete insertion during manufacture process in ensenada.Dhr lot #13907590 and relevant commodities were reviewed, the following discrepancy was found.Exception type: ncmr; document id#: (b)(4); list #: x1-5020; lot #:13854654; report quality during functional test had 2 pieces with leak.The total order quantity of (b)(4) pieces is stopped as impacted.1 sample of 2 inspected pieces was rejected and in dynamic sampling 1 sample of more than 250 inspected pieces was rejected.
|
| |
|
Event or Problem Description
|
|
The event involved a 10" (25 cm) appx 1.0 ml, smallbore ext set w/6-port nanoclave¿ manifold, check valve (orange, glow, blue, purple, yellow, green rings), nanoclave¿, luer lock where the customer reported a leak from the tubing bonding connection point.There was patient involvement but no harm or adverse event.The investigation found seal b, which makes this report 1 of 4.
|
| |
|