• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY ORANGE ENTERAL INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY ORANGE ENTERAL INFUSION PUMP Back to Search Results
Model Number INFORNG
Medical Device Problem Code Protective Measures Problem (3015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device was returned to mmd for evaluation.A dhr review was completed and did not show the currently reported issue.When the device was returned to mmdg for investigation, the pump operated as expected.Mmdg could not replicate or confirm the reported complaint.This report is being filed because there are indications that the event occurred while the device was in use by a pediatric patient (under 18) and per our procedures all reports of this type of event that involve a pediatric patient are considered reportable.
 
Event or Problem Description
The initial reporter stated that the pump was pumping after food was gone and did not alarm.They stated that they did not have any information about the pump settings.Mmd did follow up with the initial reporter who stated that they did not have any additional information about the complaint because it was not available.There were no adverse effects to the patient that were reported.Complaint:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERALITE INFINITY ORANGE ENTERAL INFUSION PUMP
Common Device Name
ENTERAL INFUSION PUMP
Manufacturer (Section D)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer (Section G)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer Contact
kristin egelhoff
4314 zevex park lane
salt lake city, UT 84123
8012641001
MDR Report Key20304735
Report Number1722139-2024-00513
Device Sequence Number7726156
Product Code LZH
UDI-Device Identifier10814844000341
UDI-Public10814844000341
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K031199
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 08/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberINFORNG
Device Catalogue NumberINFORNG
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2024
Initial Report FDA Received Date09/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/11/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-