| Catalog Number |
33650014 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Impaired Healing (2378); Metal Related Pathology (4530)
|
| Date of Event |
08/23/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Upon completion of the investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that the upon removing infinity ankle and implanting valor nail, the surgeon noticed metallosis in the surrounding tissue.Upon removal of the tibial tray there was an area of delamination visible.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction - please refer to d9 - the product was not returned for evaluation.The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that the upon removing infinity ankle and implanting valor nail, the surgeon noticed metallosis in the surrounding tissue.Upon removal of the tibial tray there was an area of delamination visible.
|
| |
|