|
It was reported that the patient arrived at the emergency department with power issues.Log file review was requested, and the event log file displayed low power advisory / power cable disconnect / low power hazard alarms while tethered on batteries.It was stated that the cause of the alarm was not identified, and that no equipment was exchanged or returned.It was stated that there was no cause found for the alarm situation.The equipment was checked in the emergency department on multiple outlets.The universal battery charger (ubc) would not power on in any outlet and there were multiple power cords tried also to confirm.The patient's ubc was replaced due to it not functioning.It was additionally stated that exchanging the ubc resolved the alarms.It was communicated on 10sep2024 that the ubc was returned because it was inoperable.The ubc did not power on due to damaged power supply assembly caused by fluid ingress.It seemed like the patient was unable to charge the batteries because they were changing depleted batteries causing controller power alarms.
|
|
Manufacturer's investigation conclusion: the reported event of alarm/functional issue with the universal battery charger (serial #(b)(6)) was confirmed with the returned unit.The unit was connected to an outlet and was not able to power on.Visual inspection revealed corrosion damage on the printed circuit boards, which is indicative of a fluid ingress issue.The customer did not provide a purchase order for the repair.The universal battery charger will be returned to the customer unrepaired and labeled ¿not for human use¿.Additional information reported that replacement of the universal battery charger resolved the alarm/functional issue.A root cause of the damaged printed circuit boards within the unit could not be determined.Heartmate 3 patient handbook rev c.Under section 3, battery charger overview, describes setting up the battery charger before use ¿before using the battery charger to charge heartmate batteries, it must be plugged in and turned on.The display panel on the front of the charger displays messages during setup and operation.¿ also, under ¿checking battery charge status using the on-battery power gauge¿, describes how to use the on-battery power gauge to confirm that the battery is fully charged.Under section 7, alarms and troubleshooting, battery related advisory messages ¿if the battery charger detects a problem with a battery, such as battery voltage too high or too low, or open battery circuit, the red light for the pocket comes on and a telephone symbol appears on the display panel to indicate a battery fault¿.Under section f, safety checklists, replace any equipment or system component that appears damaged or worn.Schedule a battery charger inspection and cleaning with thoratec-trained personnel.The safety inspection and cleaning includes (but is not limited to) functional testing, cleaning, and inspection.No further information was provided.The manufacturer is closing the file on this event.
|