ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065000133
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2024
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported that the phacoemulsification tip was bent during cataract surgery with intraocular lens implantation surgery.The surgery was completed after replacing the phacoemulsification tip with another one.There was no patient harm.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.One opened phacoemulsification tip (phaco tip), without a wrench, in a tray was received.Sample was visually inspected and found to be nonconforming.Bent at cone to cannula transition area, wear observed on threads, back of flange, or nut corners, consistent with threading on handpiece.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The exact root cause could not be determined from the investigation performed.The returned sample is visually non-conforming with the phaco tip bent; therefore, a bent phaco tip was confirmed; however, how and when the phaco tip became bent could not be determined.Most likely root cause is handling during placement of the phaco tip onto the handpiece.The exact root cause for the bent phaco tip is unknown, therefore specific action with regards to this complaint cannot be taken.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Any nonconformance, such as the phaco tip cannula bent exhibited on the returned opened sample, is removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20308431
Report Number1644019-2024-02188
Device Sequence Number12978359
Product Code LRO
UDI-Device Identifier00380650001337
UDI-Public(01)00380650001337(17)260531(11)240611(10)16VEVV
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Catalogue Number8065000133
Device Lot Number16VEVV
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/29/2024
Supplement Date Received by Manufacturer11/29/2024
Initial Report FDA Received Date09/26/2024
Supplement Report FDA Received Date12/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/11/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-