• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNOMEDICAL A/S MIO; UNO INSET II 46/6 SC1 BLUE MIMX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNOMEDICAL A/S MIO; UNO INSET II 46/6 SC1 BLUE MIMX Back to Search Results
Model Number MMT-941A
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: patient city:(b)(6).Patient country: (b)(6).Since no lot number is availabel, a detailed investigation, tests or reference samples or batch review cannot be concluded.Therefore, this complaint will be closed, this issue will be monitored through pms product trends and malfunction.If any trends picked up, this would flag on the trips and alerts according to the omqr procedure.
 
Event or Problem Description
Reference number (b)(4).Event occurred in australia.On (b)(6) 2024, it was reported that the patient encountered infusion set leaks in the tubing site.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIO
Common Device Name
UNO INSET II 46/6 SC1 BLUE MIMX
Manufacturer (Section D)
UNOMEDICAL A/S
aaholmvej 1-3
osted
lejre
Manufacturer Contact
aaholmvej 1-3
osted
lejre, 4320
MDR Report Key20308644
Report Number3003442380-2024-27306
Device Sequence Number15674360
Product Code FPA
UDI-Device Identifier05705244020009
UDI-Public05705244020009
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K160648
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-941A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2024
Initial Report FDA Received Date09/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-