|
It was reported that a retracta detachable embolization coil was difficult to detach.The device was required for embolization of a 13mm tortuous right internal iliac artery (iia) aneurysm during an abdominal aortic aneurysm (aaa) procedure.The device was stored vertically.No damage to the packaging was noted.During the procedure, the coil was advanced from the left femoral artery to the target site.Once the coil was placed at the target site, the junction zone was positioned just outside the catheter tip, and the torque device was used to turn the delivery wire 30 times; however, the coil would not detach.The delivery wire was retracted, and it was presumed that the coil had detached and was placed.The procedure continued "as usual" and was successfully completed.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.It was noted the delivery wire will be returned to ensure the coil detached correctly.A competitor¿s coils were also used during this procedure.The catheter was not flushed before each coil deployment.
|
|
This report includes information known at this time.A follow-up report will be submitted should additional, relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Correction: annex a and g.Investigation ¿ evaluation: it was reported that a retracta detachable embolization coil was difficult to detach.The device was required for embolization of a 13mm tortuous right internal iliac artery (iia) aneurysm during an abdominal aortic aneurysm (aaa) procedure.The device was stored vertically.No damage to the packaging was noted.During the procedure, the coil was advanced from the left femoral artery to the target site.Once the coil was placed at the target site, the junction zone was positioned just outside the catheter tip, and the torque device was used to turn the delivery wire 30 times; however, the coil would not detach.The delivery wire was retracted, and it was presumed that the coil had detached and was placed.The procedure continued ¿as usual¿ and was successfully completed.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.It was noted the delivery wire will be returned to ensure the coil detached correctly.A competitor¿s coils were also used during this procedure.The catheter was not flushed before each coil deployment.Reviews of the documentation, including the complaint history, device history record, manufacturing instructions and instructions for use (ifu), as well as visual inspection of the returned product, were conducted during the investigation.Cook received a used delivery wire in a stretched and elongated condition.The junction zone of the device is intact, and the coil appears to have completely deployed from the device.The junction zone that was returned is what would be expected from a device that deployed the coil without difficulty.The proximal end of the delivery wire is elongated, but this is believed to have occurred during return transit of the complaint device since this end is not involved with deployment and there was no report of elongation of the delivery wire.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) for the device found no relevant nonconformances that could have contributed to the reported failure mode.It should be noted that there were no other complaints associated with the final product lot number.Cook also reviewed product labeling: the current instructions for use [t_mwcer_rev6] state the following: "instructions for use: 1.Perform an angiogram and measure the diameter of the vessel to be occluded.6.Under fluoroscopic visualization, slowly advance the delivery wire until the entire length of the coil exits the distal end of the catheter.Ensure that the junction remains positioned just inside the catheter tip.Note: advancing the delivery wire slowly allows the junction to be seen more easily and reduces the risk of damaging it.Note: if significant resistance is encountered during coil advancement, do not continue advancing.Retract the delivery wire slightly, then gently re-advance it.If there is still significant resistance, withdraw the delivery wire from the catheter and try using a new coil with a shorter length." evidence gathered upon review of the dmr, ifu, dhr, and returned device suggests that the device was not manufactured out of specification, and that there are no nonconforming devices in house or out in the field.Based on the information provided, examination of the returned product, and the results of our investigation, it was concluded that a component failure unrelated to design or manufacturing deficiencies contributed to the reported event.Per the risk assessment no further action is required.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|