BOSTON SCIENTIFIC CORPORATION INVIVE?; CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER Back to Search Results
Model Number V173
Medical Device Problem Codes Extracardiac stimulation (1463); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code Device Overstimulation of Tissue (1991)
Date of Event 09/03/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) felt into safety mode.The patient experienced muscle stimulation, and the device was explanted and replaced.No additional adverse patient effects were reported.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) felt into safety mode.The patient experienced muscle stimulation, and the device was explanted and replaced.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode or determined it had undergone resets nd that bradycardia therapy remained available.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption (during telemetry or other higher power device operations) and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.This product was manufactured prior to implementation of the udi regulation and as a result, the complete udi is not available.Applicable us product data has been included in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVIVE?
Common Device Name
CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key20310417
Report Number2124215-2024-60334
Device Sequence Number8759290
Product Code NKE
UDI-Device Identifier00802526536632
UDI-Public(01)00802526536632
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P030005
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Remedial Action Other,Recall,Notification
Type of Report Initial,Followup
Report Date (Section B) 05/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/10/2015
Device Model NumberV173
Device Catalogue NumberV173
Device Lot Number104632
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2024
Supplement Date Received by Manufacturer05/19/2025
Initial Report FDA Received Date09/26/2024
Supplement Report FDA Received Date05/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
-
-