| Catalog Number |
08057958190 |
| Medical Device Problem Codes |
Low Test Results (2458); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We received an allegation about discrepant results for 1 patient serum/plasma sample tested with albumin gen.2 (alb2) assay and ldhi2 assay on cobas c503 analytical unit.This medwatch will apply to the ldhi2 results.Please refer to the medwatch with a1 patient identifier (b)(6) for information related to the alb2 assay.Alb2: initial result: 1.18 g/dl.Repeat result: 3.63 g/dl.Ldhi2: initial result: 380 u/l.Repeat result: 607 u/l.The physician questioned the initial result of the alb2 as it did not match the patient's clinical picture and the sample was then repeated.
|
| |
|
Additional Manufacturer Narrative
|
|
The ldhi2 reagent expiration was not provided.The alb2 reagent lot number was 786524.The expiration was not provided.The cobas c503 analytical unit serial number was (b)(6).The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The alarm trace on (b)(6)2024 was acceptable.A hardware check and precision studies were performed and they were within specifications.Based on the information provided, the investigation determined that the event was consistent with a preanalytical handling issue (it is unclear whether a 5- or 10-minute centrifugation program was used).The investigation did not identify a product problem.
|
| |
|