Catalog Number D0003
Medical Device Problem Code High Readings (2459)
Health Effect - Clinical Code Liver Damage/Dysfunction (1954)
Date of Event 08/28/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A service engineer (se) evaluated the device at the customer site.The se repeated oxygen concentrator testing several times and measurements were normal.They also checked the air pump and divertor valve function.Perfusion data was downloaded from the device and uploaded for review.The device data was reviewed by a member of the clinical team and it was determined that no issues were identified.The root cause of the reported issue could not be determined based on the information available.
 
Event or Problem Description
The device user (du) reported concerns with po2 discrepancy between the metra and their blood gas device one hour into the perfusion.While troubleshooting was being completed, it appeared that the issue might be a fluctuating po2 which gave the appearance of a difference.It was confirmed the flows were stable.The du plotted the po2 on the graphical user interface (gui) to confirm historical po2 levels which showed lots of variation.An image of the gui showed four percent o2 delivery to the liver.A manual calibration was completed which showed a small offset of 12kpa to 7kpa which the du was pleased with.However, subsequently the du reached out again to report that the instability of po2 returned.The liver was discarded because viability criteria was not met.In particular, lactate increased drastically after an initial drop before the oxygenation issue.In addition, biliary parameters worsened towards the end of perfusion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORGANOX METRA
Common Device Name
NORMOTHERMIC MACHINE PERFUSION SYSTEM FOR THE PRESERVATION OF DONOR LIVERS PRIOR
Manufacturer (Section D)
ORGANOX LIMITED
3500 john smith drive
oxford business park south
oxford, oxfordshire, gb OX4 2 WB
UK  OX4 2WB
Manufacturer (Section G)
INTEGRATED TECHNOLOGIES LTD
viking house
ellingham way
ashford, kent, gb TN23 6NF
UK   TN23 6NF
Manufacturer Contact
ruth colwill
3500 john smith drive
oxford business park south
oxford, oxfordshire, gb OX4 2-WB
UK   OX4 2WB
MDR Report Key20311276
Report Number3011560054-2024-00086
Device Sequence Number8883287
Product Code QQK
UDI-Device Identifier05060462240005
UDI-Public5060462240005
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P200035
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberD0003
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2024
Initial Report FDA Received Date09/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-