| Model Number |
15 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker contacted the customer to request additional information on the patient.Stryker requests patient information necessary for regulatory compliance, strictly adhering to hipaa's privacy rule.The customer provided stryker with all the available patient information.Patient fields in which information was not provided were intentionally left blank.Stryker evaluated the customer's device and was unable to duplicate the reported issue.Stryker cleaned the device battery fins for both battery wells and replaced the battery pins.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device unexpectedly powered off twice during a patient event.There were no reports of any adverse effects to the patient as a result of the reported issue.
|
| |
|