• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. MESHGRAFT II INSTRUMENT ONLY; EXPANDER, SURGICAL, SKIN GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. MESHGRAFT II INSTRUMENT ONLY; EXPANDER, SURGICAL, SKIN GRAFT Back to Search Results
Catalog Number 00219500100
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during kit inspection, the unit did not proper mesh.There was no patient harm/injury or surgical delay as there was no patient involvement.Due diligence is complete, no further information is available.There was no adverse event associated with this malfunction.
 
Additional Manufacturer Narrative
An investigation into the reported event has been initiated under (b)(4).Once the investigation has been completed, a follow up report will be submitted with the investigation results and any actions taken by the manufacturer.G2 country: japan.
 
Additional Manufacturer Narrative
This complaint is recorded by zimmer biomet under cmp-(b)(4).Review of the most recent repair record determined the device was not properly meshing, failed cutter.The cutter was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
There is no additional information regarding the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MESHGRAFT II INSTRUMENT ONLY
Common Device Name
EXPANDER, SURGICAL, SKIN GRAFT
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key20312056
Report Number0001526350-2024-01144
Device Sequence Number12939613
Product Code FZW
UDI-Device Identifier00889024372535
UDI-Public(01)00889024372535(11)180110(10)63898119
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
PREAMENDMENT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00219500100
Device Lot Number63898119
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2024
Supplement Date Received by Manufacturer02/05/2025
Initial Report FDA Received Date09/26/2024
Supplement Report FDA Received Date02/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/15/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RATCHET HANDLEPN: 00219502200LN AND SN: UNK
Patient SexPrefer Not To Disclose
-
-