• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORPORATION WEREWOLF FASTSEAL 6.0; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORPORATION WEREWOLF FASTSEAL 6.0; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 72290042
Medical Device Problem Codes Self-Activation or Keying (1557); Arcing (2583)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that during an unknown procedure, the werewolf fastseal wand sparked and would not power down, which meant that the device was producing plasma without being activated.The procedure was completed using a s+n back up device.There was no surgical delay.No further complications were reported.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was not received for evaluation.However, thirteen devices from the same lot number were received.A visual inspection revealed all devices are fully sealed and in their original packaging.Three devices at random were tested and no visual issues were found.A functional evaluation was performed on the three devices and found no issues activating coag in use.None of the wands got stuck being activated and none of the wands had sparks or any other issues.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEREWOLF FASTSEAL 6.0
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key20313085
Report Number3006524618-2024-00417
Device Sequence Number12787089
Product Code GEI
UDI-Device Identifier00885556787229
UDI-Public885556787229
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K210423
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72290042
Device Lot Number2149970
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2024
Supplement Date Received by Manufacturer11/01/2024
Initial Report FDA Received Date09/26/2024
Supplement Report FDA Received Date11/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-