W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
CXT201412 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Codes |
Ruptured Aneurysm (4436); Thrombosis/Thrombus (4440); Clinical Endoleak (4843)
|
| Date of Event |
07/31/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H3: code "other" was selected as the medical device remains implanted.Return not possible.H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.Product evaluation summary: the images received cannot be used to perform a full imaging evaluation because they do not meet the dicom standard.The extent and accuracy of the observations and findings may be limited due to the completeness, format and/or quality of the images provided for review.Gore cannot guarantee all key findings have been captured or that the findings are accurate.H6: d12: according to the gore® excluder® conformable aaa endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention or additional intraoperative procedure time include, but are not limited to: occlusion / stenosis of device or native vessel, endoleak, aneurysm rupture.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, the patient underwent an endovascular treatment for an abdominal aortic aneurysm using gore® excluder® conformable aaa endoprostheses and gore® excluder®aaa endoprostheses.On (b)(6) 2024, a reintervention was performed to treat a proximal type i endoleak.Two gore® viabahn® vbx balloon expandable endoprostheses (vbx) were used as chimney stentgrafts and placed in each of the left and right renal arteries, then an additional non-gore stent graft was implanted in proximal side.The endoleak was resolved, and the procedure was completed.This event was captured in medwatch report number 3007284313-2024-03449, 3007284313-2024-03450.On (b)(6) 2024, postoperative thrombotic occlusion was confirmed from the left femoral artery to those two contralateral leg endoprostheses (plc121200j, plc141400j) and the flow-divider of trunk-ipsilateral leg endoprosthesis.Thrombectomy with fogarty catheter was performed.Reportedly that the thrombosis is related to the puncture treatment.On (b)(6) 2024, the patient was transported emergency with aneurysm rupture.It was confirmed that the contralateral leg endoprosthesis (plc141400j/24704243) implanted on (b)(6) 2022 had migrated about 3 cm distally, resulting in little overlap with the contralateral gate.The rupture was considered to be due to a type iii endoleak from the overlap site.An additional stent graft was implanted below the flow-divider and the procedure was completed.Reportedly that the device may have migrated during a thrombectomy treatment performed on (b)(6) 2024.
|
| |
|
Search Alerts/Recalls
|
|
|