| Catalog Number |
MZ9270 |
| Medical Device Problem Codes |
Leak/Splash (1354); Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that bd maxzero multi-fuse extension set with needleless connector was damaged the following information was received by the initial reporter with the following verbatim occurred in june 2024: one filter leaking from the vent hole on the back of the filter.8/2/2024: second filter leaking from the vent hole on the back of the filter.
|
| |
|
Additional Manufacturer Narrative
|
|
If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer reported filter leakage, and returned one used sample of material mz5302, lot unknown.The set was infused with water, and the complaint of leakage was verified from the air vent on the filter.Supplier of filter to investigate the sample further.The supplier investigation found no issue with the filter component.The filter issue potential root cause is related to the end user drugs/solution used.A device history record review could not be performed on material mz9270 because the lot number is unknown.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|
Search Alerts/Recalls
|