Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report b5: describe event or problem g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h6: adverse event problem h10: additional narrative zimvie received one (1) tsv4h10, (imp,tsv,4.1mm,dual sel,ha) for evaluation.Visual evaluation of the as returned device identified the implant packaging / outer box was already opened.The outer vial's tamper seal / ring was broken, the inner vial and its contents were missing/not returned.The coob is non-verifiable as the condition of the product when received by the customer is unknown / non-verifiable.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.This complaint refers to the specific device being investigated for this complaint record.Device history record (dhr) review was completed for the subject lot number 1264826.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1264826 for similar events and no other complaint was identified.Review completed utilizing keywords: incorrect component quantity.Based on the investigation and risk management file review, the most likely root cause determined from the investigation is improper handling of devices/packaging outside of zimvie control.Therefore, based on the available information, a packaging malfunction could not be verified.The reported event/coob was non-verifiable since the condition of the product/packaging received by the customer was unknown / non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|