• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I CERTAIN¿ BELLATEK¿ ENCODE¿ EMERGENCE HEALING ABUTMENT 3.4MM(D) 3.8MM(P) 5MM(H); DENTAL ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I CERTAIN¿ BELLATEK¿ ENCODE¿ EMERGENCE HEALING ABUTMENT 3.4MM(D) 3.8MM(P) 5MM(H); DENTAL ABUTMENT Back to Search Results
Catalog Number IEEHA345
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).
 
Event or Problem Description
Doctor reported that the encode abutment does not seat into the implant.Procedure completed with other encode abutment.Tooth location 16.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(6).The following fields have been updated: b4: date of this report b5: describe event or problem g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h6: adverse event problem h10: additional narrative.Zimvie received one (1) ieeha345, (certain bellatek encode emergence healing abutment 3.4mm(d) 3.8mm(p) 5mm(h) for evaluation.Visual evaluation and a functional test was performed, the abutment had signs of use.It was tested in-house implant and it engaged and dis engaged as intended.Device history record (dhr) review was completed for the subject lot number 1271244.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1271244 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : fit : does not seat.A definitive root cause could not be determined since a device malfunction did not occur.The reported event was unconfirmed following functional testing and physical evaluation.Therefore, based on the available information, a device malfunction did not occur.The abutment engaged with an in-house implant as intended.The reported event was not confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN¿ BELLATEK¿ ENCODE¿ EMERGENCE HEALING ABUTMENT 3.4MM(D) 3.8MM(P) 5MM(H)
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key20320261
Report Number0001038806-2024-02192
Device Sequence Number7678765
Product Code NHA
UDI-Device Identifier00844868045403
UDI-Public(01)00844868045403(17)280822(10)1271244
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K212730
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 11/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberIEEHA345
Device Lot Number1271244
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/27/2024
Supplement Report FDA Received Date11/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/23/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
Patient EthnicityNon Hispanic
-
-