Zimvie complaint number (b)(6).The following fields have been updated: b4: date of this report b5: describe event or problem g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h6: adverse event problem h10: additional narrative.Zimvie received one (1) ieeha345, (certain bellatek encode emergence healing abutment 3.4mm(d) 3.8mm(p) 5mm(h) for evaluation.Visual evaluation and a functional test was performed, the abutment had signs of use.It was tested in-house implant and it engaged and dis engaged as intended.Device history record (dhr) review was completed for the subject lot number 1271244.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1271244 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : fit : does not seat.A definitive root cause could not be determined since a device malfunction did not occur.The reported event was unconfirmed following functional testing and physical evaluation.Therefore, based on the available information, a device malfunction did not occur.The abutment engaged with an in-house implant as intended.The reported event was not confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|