• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PRLNE HS BLU 24IN 5-0 D/A BV-1; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PRLNE HS BLU 24IN 5-0 D/A BV-1; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Catalog Number 9702H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure in 2024 and suture was used.During the procedure, the proline suture, needle breaking off.There were no adverse consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Were there any patient consequences? if yes, please describe.Please provide lot number.Please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep).Please confirm if the devices are available for product return.If yes, please provide the status of the device(s) as it has not been received for analysis.If the device has been shipped, please provide the shipment tracking details.The manufacturing records could not be reviewed as the batch/lot number is unknown.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees, that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional h6 component code: g07002 no device problem found.Investigation summary: the product was returned to ethicon for evaluation.One unused opened sample was received for analysis.Product code 9702h.In order to evaluate the conditions of the returned sample, the swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along the strand no anomalies or pull off was observed during the evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional investigation summary: the product was returned to ethicon for evaluation.One open box with thirty-three representative unopened samples was received for analysis.Product code 9702h.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damages were observed on the external packets.As per the sampling plan, a visual inspection was performed on thirteen samples.The packets were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE HS BLU 24IN 5-0 D/A BV-1
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key20321615
Report Number2210968-2024-10354
Device Sequence Number20809124
Product Code GAW
UDI-Device Identifier10705031022867
UDI-Public(01)10705031022867(17)290430(10)101AQZ
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number9702H
Device Lot Number101AQZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2024
Supplement Date Received by Manufacturer11/15/2024
Initial Report FDA Received Date09/27/2024
Supplement Report FDA Received Date12/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-