TORNIER S.A.S. AEQUALIS REVERSED GLENOSPHERE LAT.4 42MM X 29MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number DWD194
Medical Device Problem Codes Unstable (1667); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); Joint Laxity (4526); Muscle/Tendon Damage (4532)
Date of Event 09/02/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that there was a need for a revision of reverse shoulder due to dislocation following primary reverse.The surgeon decided to revise the implants in place to give greater lateralization.He also used a thicker constrained liner.The patient had a torn subscapularis and therefore the joint was unstable which required this revision.Implants were revised to give greater stability and more lateralization.
 
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.
 
Additional Manufacturer Narrative
Corrected fields: h6 - device code grid, clinical signs code grid, and health impact code grid.The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided the investigation report will be reassessed.
 
Event or Problem Description
It was reported that there was a need for a revision of reverse shoulder due to dislocation following primary reverse.The surgeon decided to revise the implants in place to give greater lateralization.He also used a thicker constrained liner.The patient had a torn subscapularis and therefore the joint was unstable which required this revision.Implants were revised to give greater stability and more lateralization.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AEQUALIS REVERSED GLENOSPHERE LAT.4 42MM X 29MM
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR   38330
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key20321944
Report Number3000931034-2024-00586
Device Sequence Number7630566
Product Code KWS
UDI-Device Identifier03700386955136
UDI-Public(01)03700386955136(11)240427(17)290427(10)6666BA010
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K081059
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberDWD194
Device Lot Number5228AY
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/02/2024
Supplement Date Received by Manufacturer11/28/2024
Initial Report FDA Received Date09/27/2024
Supplement Report FDA Received Date12/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexMale
-
-