MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW CVC SET: 16 GA X 8" (20 CM); CATHETER,INTRAVASCULAR,THERAP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW CVC SET: 16 GA X 8" (20 CM); CATHETER,INTRAVASCULAR,THERAP Back to Search Results
Catalog Number CS-24301-E
Medical Device Problem Codes Material Twisted/Bent (2981); Material Too Soft/Flexible (4007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported "on (b)(6) 2024,the swg is soft, it cannot be used together with the dilator when used on the patient.No harm for the patient." additional information reports "the swg is easy to kink." the user changed to a new set.There was no medical intervention required.The patient's current condition is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).The customer returned one opened cvc kit for analysis.No definite signs of use were observed.The reported guide wire was not returned for analysis.No defects or anomalies were observed with the other returned components.A device history record review was performed and no relevant findings were identified.The customer report of a kinked guide wire could not be confirmed through investigation of the returned sample.The guide wire was not returned; therefore, relevant inspections could not be performed.A device history record review revealed no relevant findings.Based on the customer report and without the guide wire returned for analysis, the potential root cause cannot be confirmed at this time.Teleflex will continue to monitor and trend on reports of this nature.
 
Event or Problem Description
It was reported "on (b)(6) 2024,the swg is soft, it cannot be used together with the dilator when used on the patient.No harm for the patient." additional information reports "the swg is easy to kink." the user changed to a new set.There was no medical intervention required.The patient's current condition is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 16 GA X 8" (20 CM)
Common Device Name
CATHETER,INTRAVASCULAR,THERAP
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key20322202
Report Number3006425876-2024-00986
Device Sequence Number12708459
Product Code FOZ
UDI-Device Identifier00801902102225
UDI-Public(01)00801902102225(11)231201(17)251018(10)71F23L1401(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K900263
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date10/18/2025
Device Catalogue NumberCS-24301-E
Device Lot Number71F23L1401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2024
Supplement Date Received by Manufacturer10/04/2024
Initial Report FDA Received Date09/27/2024
Supplement Report FDA Received Date10/09/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED.; NONE REPORTED.
Patient SexUnknown
-
-