MAUDE Adverse Event Report: BECTON DICKINSON & CO. (SPARKS) BD PHOENIX¿ PMIC/ID-106; SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON & CO. (SPARKS) BD PHOENIX¿ PMIC/ID-106; SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION Back to Search Results
Catalog Number 448606
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There were multiple 510k numbers reported to be involved.The information for the additional 510k is as follows: g4.Pma / 510(k)#: k020322, k021954, k022172, k023273, k023301, k024152, k030677, k031306, k031679, k031679, k032131, k033784, k033907, k040006, k040106, k040716, k050555, k051689, k053241, k060214, k060217, k060218, k060218, k060493, k082538, k082852, k082913, and k131331.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using the bd phoenix¿ pmic/id-106 the user experienced misidentifications with staphylococcus species from a patient wound specimen.No health impact or consequence reported.Report 2 of 2.
 
Additional Manufacturer Narrative
Investigation summary: this complaint is for misidentification of staphylococcus when using phoenix panel pmic/id-106 (catalog number 448606) batch number 3353080.The customer did not return panels or isolates but provided phoenix generated lab reports and photos for the investigation.The customer provided phoenix lab reports show an isolate identified as staphylococcus epidermidis.The photos show the sequence numbers and dates.To investigate, retention panels from the complaint batch were tested using in house and qc isolates s.Aureus a25923, s.Aureus a29213, s.Aureus 4757 and s.Aureus a43300 on a phoenix m50 machine and evaluated for identification results.Additionally, one control panel from the same material but different batch were inoculated with qc isolate s.Aureus a29213 on a phoenix m50 machine and evaluated for identification results.The panels tested identified the isolates as s.Aureus, therefore, this complaint is not confirmed for misidentification.The batch history record was satisfactory, and no quality notifications were generated during manufacturing and inspection.Complaint trending was performed, and no trends were identified associated with this defect.Bd will continue to monitor for trends and take action as necessary.
 
Event or Problem Description
It was reported while using the bd phoenix¿ pmic/id-106 the user experienced misidentifications with staphylococcus species from a patient wound specimen.No health impact or consequence reported.Report 2 of 2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHOENIX¿ PMIC/ID-106
Common Device Name
SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION
Manufacturer (Section D)
BECTON DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key20322748
Report Number1119779-2024-00694
Device Sequence Number12763990
Product Code LON
UDI-Device Identifier30382904486069
UDI-Public(01)30382904486069(11)231219(17)241218(10)3353080
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
SEE H11
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/18/2024
Device Catalogue Number448606
Device Lot Number3353080
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2024
Supplement Date Received by Manufacturer10/30/2024
Initial Report FDA Received Date09/27/2024
Supplement Report FDA Received Date11/19/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-