|
MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS; EXPANDER, SKIN, INFLATABLE
|
Back to Search Results |
|
| Catalog Number |
3549213 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the product was returned to mentor for evaluation.Mentor then conducted a visual inspection, leak testing, and microscopic examination of the returned device.During the visual analysis of the returned sample, no apparent damage or visual anomalies were observed.Leak testing was performed, according to the mentor procedure, and a tear was found in the shell on the anterior view, measuring approximately 0.2 cm.No other leak sites were found during the analysis.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in the whole area of the tear.Parallel striations are consistent with markings made by a sharp object perforating the tissue expander shell.Based on the information currently available, the microscopic examination of the returned product indicates that the expander could have been damaged during the implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps, or scissors to contact the device during the implantation.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a 450cc mentor cpx 4 breast tissue expander with suture tabs was found to be deflated during an operation.No adverse event was experienced by the patient.The surgeon used backup devices, and there were no reported patient consequences or procedural delays.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|