|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.Device was returned for evaluation and the customer's allegation was confirmed.The device evaluation found that the probe was broken at 14.2 mm from its distal end.Based on the results of the investigation, a definitive root cause cannot be identified, however, it is likely that the following led to the malfunction: 1) during the output activation in seal & cut mode, the probe was contacting hard tissue, metal objects or surgical instruments.2) due to ultrasonic vibration, the coating of the probe peeled off.Also, scratches were generated.3) a force to activate the output in seal & cut mode, or a force to grasp the tissue was applied to the probe.Therefore, cracks were generated at the scratched area.4) a force was applied to the probe causing it to break.A device history review revealed no issues that could have caused or contributed to the reported issue.Labeling: this issue is addressed in the instructions for use (ifu): drawing number and revision number of ifu: rc2994 12 - the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator, forceps, and others).Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/splitting/protruding/partial separating of the tissue pad.In turn, the probe may break before displaying an error window or generating an alarm tone.- do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities.Olympus will continue to monitor the field performance of this device.
|