| Catalog Number |
33680001 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/30/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is not available for evaluation as it remains implanted in the patient.A review of the device history is not possible because the lot number was not communicated.If additional information becomes available, it will be provided on a supplemental report.
|
| |
|
Event or Problem Description
|
|
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery for reasons that are not available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Failure of total ankle replacement tar over time is a known complication.In case of a planned revision procedure a medical opinion aiming to determine the root cause of the failure is part of the internal investigation process.Sufficient radiological and clinical information must be provided to enable a meaningful clinical assessment and to identify possible causes of failure.In cases where a ct scan is the only available clinical source, this assessment is limited.No conclusive statement can be provided, because key clinical information e.G.Clinical status, exact symptoms and range of motion of the patient, assessment of the treating physician is missing.The root causes of radiographic findings such as radiolucent areas and bone cysts are often complex and multifactorial and cannot be determined scientifically without this decisive evidence.Due to these limitations, we are unable in such cases to provide statements about the patient, the procedure and or the device in relation to cause of the failure.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery for reasons that are not available at the time of this report.
|
| |
|
Search Alerts/Recalls
|