Catalog Number 05.001.231
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 09/04/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: investigation summary: a video was provided by the customer for evaluation.Based on the provided video, it was determined that the device had the following failures: unexpected disengagement - battery/case/lid and deformed/bent - housing.Therefore, the reported condition was confirmed.However, an assignable could not be determined.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: h4: the lot number was unknown; therefore, the device manufacture date is unknown.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event or Problem Description
This is report 2 of 2 for (b)(4).It was reported from colombia that during an unspecified surgical procedure, a battery handpiece device and a lid for handpiece device were used together.According to the report, the devices were tested at the beginning of the procedure before incising the patient¿s skin, but they were not able to close the lid device on the battery handpiece device.It was reported that the surgical team was already sterile, and the patient was already under spinal anesthesia when the event occurred.It was reported that they tried using another handpiece device from a competitor¿s brand but it did not fit the insertion system for the drills.It was reported that although an attempt was made to solve this problem with the closure of the handpiece device with the lid device, they decided to cancel the procedure.The status of the patient was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LID-HANDPIECE/MODULAR FOR TRS
Common Device Name
BATTERY, REPLACEMENT, RECHARGEABLE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key20327543
Report Number8030965-2024-12024
Device Sequence Number12728431
Product Code MOQ
UDI-Device Identifier10886982130143
UDI-Public10886982130143
Combination Product (Y/N)N
Initial Reporter CountryCO
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05.001.231
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2024
Supplement Date Received by Manufacturer10/29/2024
Initial Report FDA Received Date09/27/2024
Supplement Report FDA Received Date10/29/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-