| Model Number |
A22005A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported, the telescope, 30°, wideangle, 4 mm exhibited a broken ceramic tip.The issue occurred during preparation for use for a therapeutic hysteroscopy procedure.The procedure was completed using a similar device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Additional information: d8, d9, h3, h4, h6, h11 the device was returned to olympus for inspection, and the reportable malfunction was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the reported event is most likely due to the effect of a large mechanical force due to a shock, fall or collision with other equipment.Olympus will continue to monitor field performance for this device.
|
| |
|