| Model Number |
429688 |
| Medical Device Problem Codes |
Failure to Capture (1081); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient exhibited heart rates in the 40s.It was observed that the left ventricular (lv) lead exhibited high thresholds and a possible loss of capture.Additionally, the right atrial (ra) lead exhibited undersensing on current and stored e lectrograms (egm), possibly resulting in false termination of atrial tachycardia/atrial fibrillation (at/af) recorded episodes.The implantable pulse generator (ipg) system was inactivated and replaced.No patient complications have been reported as a result of th is event.
|
| |
|