| Model Number |
8253001RF |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Inappropriate Audible Prompt/Feedback (2280)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during thyroidectomy procedure, prior to incision but after intubation, the surgeon stimmed the patients neck and got a normal response.During the procedure, a trivantage 2 was utilized.However, when the surgeon attempted stimulation, the machine made a tweedled noise, indicating no stim.Surgeon stimmed the vagus nerve inside the carotid sheath with 3 milliamps and got no visual movement in the muscle.There was around 1 hour delay in the procedure.The surgery was aborted on the contralateral side.The tube was working properly during pre-op.The issue was not resolved by repositioning or troubleshooting.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: code is updated.Previously updated fdc d16 is no longer available.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|