Model Number MMT-1780K
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Unit passed displacement test and active current measurement.Unit received with unexpected battery power loss in power graph and had high sleep current, problem isolated to pcb 2.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported to medtronic minimed that the customer reported unexpected battery power loss.The customer reported no adverse event.The event involved product mmt-1780k.Troubleshooting was performed and the pump will be replaced.No harm requiring medical intervention was reported.The customer will discontinue the use of the device.Mmt-1780k was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP MMT-1780K 670G 3ML BLACK
Common Device Name
AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
lusine boyadzhyan
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
MDR Report Key20331200
Report Number2032227-2024-246230
Device Sequence Number13168906
Product Code OZP
UDI-Device Identifier00643169939202
UDI-Public(01)00643169939202
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P150001
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 09/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-1780K
Device Catalogue NumberMMT-1780K
Device Lot NumberHG1XCDA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2019
Initial Report FDA Received Date09/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/28/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2032227-07-23-2024-001-C
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
Patient Weight92 KG
-
-