| Model Number |
T10 DRIVER, RETAINING |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2024, it was reported by a sales representative via sems-06666422 that (b)(4)) of a 5042-100 t10 driver was used with a torque limiting handle, yet the threads were bent, so they did not properly grip the screws.No pieces broke inside the patient.The case was completed successfully using a t10 driver from the 2.7 mini frag set without issues.This was discovered during a procedure on (b)(6) 2024, with no reported patient harm.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.The most likely cause for the reported failure can be attributed to overstressing a device that is damaged naturally and inevitably as a result of normal wear or aging from repeated usage/reprocessing.
|
| |
|