• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US LCS MENIS BEARING STD+/10MM; LCS : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US LCS MENIS BEARING STD+/10MM; LCS : KNEE TIBIAL INSERT Back to Search Results
Catalog Number 117824025
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/01/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Revision of an lcs meniscal bearing in patient x necessary due to wear of the insert.Revision is not planned yet.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, "revision of an lcs meniscal bearing in patient x necessary due to wear of the insert.Revision is not planned yet".The product was not returned to j&j medtech orthopaedics, however photos were provided for review.See attachment (product info).Photos were provided for review, however the photos only show part and lot number and no other observation pertaining to the nature of the reported event could be identified.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was not confirmed as the photographs provided contained insufficient evidence of the reported event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LCS MENIS BEARING STD+/10MM
Common Device Name
LCS : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key20332605
Report Number1818910-2024-20399
Device Sequence Number18167036
Product Code NJL
UDI-Device Identifier10603295007807
UDI-Public(01)10603295007807(17)190331(10)7879105
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2019
Device Catalogue Number117824025
Device Lot Number7879105
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/16/2024
Supplement Date Received by Manufacturer11/11/2024
Initial Report FDA Received Date09/30/2024
Supplement Report FDA Received Date11/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/17/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-