| Model Number |
320749 |
| Medical Device Problem Code |
Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Consumer reported finding the pen needle would clog during the flow check.Consumer reported finding the pen needle would work during the flow check but clogged during the injection.Dc.Lot # 3325695; catalog# 320749; date of event unknown; sample status discard.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: no samples (including photos) were returned; therefore, the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
|
| |
|