Visual inspection: during the investigation, no significant deformations or damage of the valve were determined.Permeability test: a permeability test has shown that all components are permeable.Computer controlled test: to investigate the valve, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in the horizontal position.The results show that the progav 2.0 does not operate within the permissible tolerance in the horizontal position.An accelerated could be determined.Adjustment test: the progav 2.0 was tested and is not adjustable.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected; however, the breaking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valve, deposits were found inside.Results: in advance, we would like to point out that the product sent in was not inserted in liquid at the time of delivery.Dried deposits can influence the function of the products, which can affect the results.Despite this, we have examined the valve.Based on our investigation results, we can determine an accelerated outflow and a non-adjustability at the progav 2.0.The visible deposits in the valve have led to the functional impairment.Deposits caused by substances naturally present in the patient's body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic material can affect the integrity of the valve.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|