|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Additional information: a2, a4, a5, a6, b5, b6, and b7.The device was returned to the repair base for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: there was water ingress in the imaging section and in the cable, there was corrosion of electrical contacts, a loose scope mount, and a crack in the main body.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: regarding the customer¿s allegation and for all the newly reportable malfunctions mentioned above the cause was traced to component failure.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|